The ac-SECM is a new technique that has the capability to measure surface phenomena without the presence of a mediator. This feature allows the SECM technique to be used in the field of corrosion science.
The SECM technique can be combined with the OSP470 module or the ic-SECM470 module to allow the technique to make both constant height and constant distance measurements.
Potentiostat
- Compliance voltage: ±12 V
- Applied potential: ±10 V FSR
- Resolution: 32-bit (4.7 nV)
- Measured potential: ±10 V FSR
- Resolution: 24-bit (1.2 μV)
- Current ranges: 10-decades 1 nA to 1 A
- Maximum current: 500 mA
- Current resolution: 23.8 fA
- Accuracy: >0.5%
- Floating capability Standard
- Cell connections 2, 3 or 4
- Maximum ADC sample rate: 4 MHz
- Maximum ADC resolution: 24-bit
- Minimum pulse duration: 100 μs
- Scan rate: 1 μV/s to 200 V/s
Electrometer
- Impedance: 10¹³ Ω II 7 pF typical
- Bias current: 1 pA typical
- Bandwidth 1 MHz
EIS Capability
- Frequency range: 1 µHz to 1 MHz
- Analyser accuracy: 0.1%, 0.1°
- Frequency resolution 66 nHz
- Mode Single sine, Multisine, FFT analysis